Set Cal Timers

5-14 | Thermox Series 2000 WDG Insitu Analyzer
Figure 5-2. Set Cal Timers menu options.
Sub-Menu
Set Auto Timers
Set Time and Date
Auto Cal On/Off
Cal Gas Duration
Recovery Duration

Set Cal Timers - Set Auto Timers

Set the frequencies for automatic calibrations or verifications, including
the time when the first calibration or verification should begin. You have
the option of setting the cycle in either hours or days.
If you schedule a verification and calibration at the same time, the
verification will be performed first.
NOTE
SET CAL TIMERS KEY
Set Auto Timers
Set Time & Date
Aut Cal On/Off
Cal Gas Duration
Recovery Duration
Function
Set auto timer for desired auto calibration/verification
cycle times.
Set the current time and date.
Enable or disable the auto calibration cycle.
Define how long cal gases should flow through the sen-
sor during automatic calibrations or verifications.
Define a recovery duration during which time the sensor
switches from measuring calibration gas readings to
measuring process readings without affecting the ana-
log output or triggering of alarms.