Set Cal Timers - Set Time And Date - Ametek Thermox WDG Insitu User Manual

With series 2000 control unit
Table of Contents

Advertisement

1.
From the Set Cal Timers menu, select Set Auto Timers.
2.
Select Auto Cal Cycle if defining an automatic calibration cycle. Select
Auto Verify Cycle if defining an automatic verification cycle.
3.
Select Set Cycle in DAYS or Set Cycle in HOURS, depending on
whether you want to set the frequency of calibrations or verifications
in days or hours.
If you are setting the auto timers for the first time, the Series 2000
software will automatically prompt you to enter the current time and
date.
NOTE
You will be prompted to enter the frequency for the calibration or verifica-
tion cycle (either in days or hours, depending on what you selected:
4.
Enter the number of days (or hours) for the automatic calibration or
verify cycle. As you enter the values, they overwrite the XXs on the
display. To enter 1 day or 1 hour, enter 01; otherwise, the system will
assume you are entering 10 days (or hours).
You will be prompted to enter the time of day when the first calibra-
tion of the cycle should begin:
5.
Enter the time you want the first calibration or verification to begin,
then press Enter. The system uses military time, so to enter 4:00 p.m.,
you would type 16:00. If, for example, you set the cycle frequency at
8 hours, and specify the first calibration to take place at 4:00 p.m., the
next automatic calibration or verification will take place at midnight.

Set Cal Timers - Set Time and Date

Cal Cycle = 01 days
New Value = xx
Cal time is YY:YY
Enter new time Hr:Mn
Calibration | 5-15