General Wiring And Conduit Requirements - Ametek Thermox WDG Insitu User Manual

With series 2000 control unit
Table of Contents

Advertisement

This wiring section shows you how to make the following connections:
AC mains supply wiring to control unit
AC mains supply wiring to sensor
Control unit to sensor
Control unit to remote calibration unit
Control unit to alarm devices
Control unit to current output devices
Control unit to host computer (RS-485 communications)
In addition, this wiring section provides mandatory EMC grounding, shielding,
and noise protection requirements.
If you are considering adding option cards at a later time, such as a
combustibles card, you may want to connect the required intercon-
necting wires when you first install your analyzer, especially if run-
NOTE
ning the wires over long distances through conduit.

General Wiring and Conduit Requirements

This section describes general wiring and conduit requirements.
Sensor wiring conductors must be rated at a minimum of 80 °C. All other
wiring conductor ratings should be for the minimum temperature required for
the equipment being connected to the analyzer, but not less than 60 °C.
Use only the applicable NEMA-approved conduit fittings or cable fittings to
maintain the NEMA rating for the Series 2000 controller or sensor enclosure.
If not using a conduit entry, leave the factory NEMA-approved plugs intact.
Never leave any holes unplugged.
Follow all applicable electrical codes for your location.
Follow proper grounding, shielding and noise protection practices as de-
scribed in this section.
Use twisted-pair cable with an overall braided shield, or twisted-pair cable in
rigid metal conduit for all sensor and signal wiring.
Installation and Start-Up | 3-13