Overview; Aborting A Calibration - Ametek Thermox WDG Insitu User Manual

With series 2000 control unit
Table of Contents

Advertisement

Overview

Aborting a Calibration

5-4 | Thermox Series 2000 WDG Insitu Analyzer
CALIBRATE KEY
Menu Options
Initiate Cal/Initiate Verify
Although automatic calibrations can be set to run at predetermined frequencies,
you can also force the system into an automatic calibration.
Cal/Verify Data
Cal Gas Values
Inject Cal Gas
Set Cal Timers/Set Auto Timers
(applies only for auto cal operations
with RCU, except for setting time and
date)
Pressing the Calibrate key while the system is performing a calibration
or verification, displays the option to abort the calibration or continue the
calibration. All other menu functions are temporarily disabled during a
calibration or verification. Oxygen, combustibles, and methane alarms are
also disabled during a calibration.
Functions/Options
Start an automatic, remote, or manual cali-
bration or verify
View latest calibration and verify data.
- Cal data
- Verify data
Define calibration/verify gas values to
match known values of calibration gas
cylinders.
Enter calibration gas values.
Set pressure and flow to RCU gases with-
out performing a calibration or verification.
Set control unit time/date.
Enable or disable auto cal cycles.
Set cal gas times for auto cal/verify opera-
tions.
Define recovery durations after a cal/verify.