General Procedures and Functional Checks
Power shut down (continued)
NOTE:
System shutdown
CAUTION
4-12
The SYSTEM - EXIT menu, used when switching off the unit,
gives you these choices:
Logoff
Use this button to log off the current user.
The system remains ON and ready for a new user to log on.
If the Logoff button is dimmed, it indicates that no user is
logged on to the unit at the moment.
Shutdown
Use this button to shut down the system. The entire system
will shut down. It is recommended to perform a full shutdown
at least once a week.
Sleep Mode
Use this button when you do a portable exam in order to
reduce the time to start up the system.
Cancel
Use this button to exit from the System-Exit menu and
return to the previous operation.
Exit
(Only available when logged in as GE Service with Service
Dongle)
Select this button when you want to exit to the Windows
Desktop.
If you need to restart LOGIQ e when logged on to the
Windows Desktop, ensure that you do a complete power
down (Shut Down). This is required to power up the Front
End Processor.
Disconnect the Mains Power Cable is necessary. For example:
Relocating the scanner.
DO NOT unplug and/or transport the unit until after the power
off sequence has been completed. Failure to do so may result
in corrupted patient files.
LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6