Electrical Requirements - GE LOGIQ e Basic Service Manual

Hide thumbs Also See for LOGIQ e:
Table of Contents

Advertisement

Site Preparations

Electrical requirements

General requirements
NOTE:
NOTE:
Electrical requirements for Adapter
2-6
GE requires a dedicated power and ground for the proper
operation of its Ultrasound equipment. This dedicated power
shall originate at the last distribution panel before the Ultrasound
system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not
shared with any other circuit), and a full size ground wire from
the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not
shared with any other circuit, and a full size ground wire from the
distribution panel to the Ultrasound outlet.
Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power
source to the Ultrasound system is only a conduit.
Table 2-5:
Adapter Model
Input Voltage
Input Current
Output Power
Output Voltage
Output Current
Frequency
Electrical Specifications for Adapter
AHM150PS19-XA1048
100-240VAC
1.5A at 115VAC
0.75A at 230VAC
150W
19.0VDC
7.89A
50/60Hz
LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6