Step
13
Change scale (PRF)
14
Transparency Map
15
Focus Position
16
Power output
17
Wall Filter
18
Map Compress
19
Accumulation
20
Sample Vol
21
Flash Suppression
LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6
Table 4-6:
Color Flow Mode Softmenu Key
Task
Velocity scale determines pulse repetition
frequency. If the sample volume gate range
exceeds single gate PRF capability, the
system automatically switches to high PRF
mode. Multiple gates appear, and HPRF is
indicated on the display.
Allows to select specific transparency map
Increases the number of transmit focal zones
or moves the focal zone(s) so that you can
tighten up the beam for a specific area. A
graphic caret corresponding to the focal zone
position(s) appears on the right edge of the
image.
Optimizes image quality and allows user to
reduce beam intensity. 10% increments
between 0-100%. Values greater than 0.1 are
displayed.
Wall Filter insulates the Doppler signal from
excessive noise caused from vessel
movement.
When you increase the value, high velocity
elements in the map are compressed so that
the map darkens. When you decrease the
value, low velocity elements in the map are
compressed so that the map lightens.
Accumulation enhances the flow in an image.
Off-Infinite: Infinite provides the same result as
applying CINE Capture to a B-Flow CINE clip.
Adjust the control to adjust the sample volume
gate on the Color Flow image.
Activates/deactivates Flash Suppression, a
motion artifact elimination process.
Expected Result(s)
Functional checks
Comments
4-31