General Procedures and Functional Checks
Connect AC/DC to the LOGIQ e
4-6
Connecting AC/DC to the LOGIQ e ultrasound unit, involves
preliminary checks of the power cord, voltage level and
compliance with electrical safety requirements.
1. Ensure that the wall outlet is of appropriate type.
2. Uncoil the power cable, allowing sufficient slack so that the
unit can be moved slightly.
3. Verify that the power cable is without any visible scratches
or any sign of damage.
4. Verify that the on-site mains voltage is within the limits
indicated on the rating label near the Circuit Breaker on the
rear of the unit.
5. Connect the Power Cable to the Power Inlet at the rear of
the unit.
6. Connect the Power Cable's other end (male plug) to a
hospital grade mains power outlet with the proper rated
voltage, and the unit is ready for Power ON/Boot Up.
LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6