CH1
CH2
 ③ In case large waveform-noise (as shown in Fig. 1) is detected, use the FILTER function.
  �ow to set FILTER function (e.g. DL1540 manufactured by YOKOGAWA)
  1. Press the FILTER button.
  2. Select "Smooth" of the menu on screen and turn it ON.
④ In case the amplitude is small, loosen the two screws (#541)
that attach the GMR unit, and adjust the mounting angle of
the GMR unit.
⑤ In case the adjustment cannot be made in ④ , check the
MR head. If it is deformed, make a bending correction
for the MR head. If the correction is impossible or there
is no deform, replace the GMR unit. (Fig. 2)
Note: When adjustments are made, prevent the magnetic
surface and MR head from touching the magnetized
driver bit. Otherwise, the magnetic data may be
damaged.
Fig.1
- A 3・AF-S VR 70-300/4.5-5.6G -
JAA79551-R.3690.A
● Oscilloscope setting
 V/Div (ch1)
 V/Div (ch2)
 Coupling
 Time/Div
 Trigger Mode
 Trigger Coupling
Amplitude
Standard: Amplitude of all pulses/wave-
forms is 80mV or more.
Note: Check the waveform by moving the
focus ring back and forth from the
infinity-end to the close-end positions
entirely.
MR head
Fig.2
INC
: 50 mV
: 50 mV
: AC
: 2 m Sec
: NORMAL
: AC
#541×2
GMR unit