Equipment Overview: DINAMAP® Models PRO Series 100-400 and ProCare
the external monitor, determine the baudrate setting for the 250 Series monitor.
Then exit the service mode screens.
3.
Access the General Setup screen and choose the time interval for printing the
MSpO
2
detailed information.
4.
Ensure the NPB N-200 is set to the Beat-to-Beat mode.
Through this interface, FSpO
printed at selected intervals on the strip chart paper of the 250 Series Monitor.
When the recorder is on, each reading is printed in the annotation area between the
top and bottom grids. A filled diamond marker, above the measured values, marks
the time of the reading and identifies the data source as an external device. The
following is an example annotation:
FSpO
45%
2
In addition, the FSpO
chart paper; the trend is printed as a beaded trace (
NOTE: Your Nellcor N-400 must be at software version 4.2.0.02 or later. Consult
your Nellcor product documentation, or contact your Nellcor
representative if you are not certain of the version.
To connect the Model N-400:
1.
Using interface cable, connect one end to an available
J110, or J111) on the 250 Series Monitor; connect the other end to the
Communications
Access the Communications Setup service mode screen and set the mode to Nellcor
2.
for the appropriate port. Using the manufacturer's instructions for use for the
external monitor, determine the baudrate setting for the 250 Series monitor. Then
exit the service mode screens.
Access the General Setup screen, then: choose the time interval for printing the
3.
FSpO
values on the strip chart paper; and enable/disable trending of FSpO
2
the bottom grid of the strip chart paper. Refer to
information.
4.
Ensure the NPB N-400 is set to the Beat-to-Beat mode.
All of the above DINAMAP Monitors can be interfaced to a 250 Series Monitor to
provide a printout of NIBP values on the strip chart paper.
250 Series Maternal/Fetal Monitor
values on the strip chart paper. Refer to
readings provided by an NPB Model N-400 are
2
trend can be optionally printed on the bottom grid of the strip
2
port on the pulse oximeter.
2020551-001
"Setup"
on page 3-9 for more
).
port (J109,
RS-232C
Serial
2
"Setup"
on page 3-9 for more
2-15