Making the Electrical Connections
[1]
Determine the customer's electrical service. See the table below.
* L1, L2, and Neutral used in this configuration are sometimes referred to as Single Phase Connections.
16
Connecting the Processor
Dangerous Voltage.
Service Options
Voltage
Frequency
volts
Hz
200
50/60
220
50/60
230
50/60
240
50/60
100/200
50/60
120/240
50/60
120/208
60
127/220
50
220/380
50
230/400
50
240/415
50
200
50/60
120/208
60
127/220
50
220/380
50
230/400
50
240/415
50
Service
2-wire, single phase
2-wire, single phase
2-wire, single phase
2-wire, single phase
3-wire, single phase
3-wire, single phase
3-wire, 3-phase*, wye
3-wire, 3-phase*, wye
3-wire, 3-phase*, wye
3-wire, 3-phase*, wye
3-wire, 3-phase*, wye
3-wire, 3-phase, delta
4-wire, 3-phase, wye
4-wire, 3-phase, wye
4-wire, 3-phase, wye
4-wire, 3-phase, wye
4-wire, 3-phase, wye
NOTE
2B6843