P
OWER
The Gemini VII 2390 is designed to operate with a universal input power supply (100-240VAC) at
50/60Hz. Noise-free power of the correct voltage and frequency, with a safety earth ground,
should be available through a standard wall receptacle. The power outlet should be able to supply
15 amps @ 100 or 110VAC ±10% or 7.5 amps @ 230VAC ±10%. These requirements can be
checked by using a voltage meter (available at most hardware or electronic supply houses) or a
multimeter. There should also be sufficient outlets for the computer, monitor, printer, and any
other peripheral devices.
The analyzer and peripheral devices must be installed on their own dedicated power
line. Other devices — such as motors, generators, or ovens — should not be placed
on the same power line.
Replacement power supply cords must be rated for the specifications stated above.
P
A
A
ARTS
ND
CCESSORIES
Parts and accessories can be found online at www.Micromeritics.com.
Gemini VII 2390 Installation Instructions and Checklist
239-42870-02 Rev (B) — May 2021