The AccuPyc III uses a 24 V, 11.7 amp power supply with an input range of 100-240 VAC (±10%),
280 VA, 50-60 Hz. Noise-free power of the correct voltage and frequency, with a safety earth
ground, should be available through a standard wall receptacle. There should also be sufficient
outlets and easy access to those outlets for all devices.
The external power adapter required for the AccuPyc III is Micromeritics' part number
003-40001-02. Use of any other power adapter could damage equipment and/or
cause harm to the operator. The AccuPyc III is intended to be powered from the
output of the approved power adapter rated Class I, manufactured by Mean Well, P/N
GST280A24-C6P. Micromeritics supplies a suitably rated approved power supply
cord appropriate for the applicable country with the power adapter.
The analyzer and peripheral devices must be installed on their own dedicated power
line. Other devices — such as motors, generators, or ovens — should not be placed
on the same power line.
This instrument does not have a power switch and must be disconnected from the
wall outlet when powering off; therefore it is important to position the instrument where
the wall outlet is easily accessible.
T
EMPERATURE AND
Temperature and humidity must be controlled to within:
Temperature: Stable between 10 to 35 °C (50 to 96 °F). Temperature change up to 2 °C per
hour.
Humidity:
10 to 80% relative (non-condensing) for laboratory temperature from 10 to 25 °C,
maximum RH decreasing linearly to 50% at 35 °C.
Do Not:
Allow room temperature or humidity to exceed limits.
n
Install the analyzer where it is exposed to direct sunlight.
n
Locate the analyzer near air conditioning or heating vents.
n