Software Installation In 21Cfr11 Environments - Micromeritics GEMINI VII 2390 Installation Instructions And Checklist

Surface area analyzer
Hide thumbs Also See for GEMINI VII 2390:
Table of Contents

Advertisement

S
I
OFTWARE
NSTALLATION IN
The following Confirm application prerequisites must be met prior to installing the software:
Windows 10 Professional or Windows 10 Enterprise (or higher) must be installed.
n
Configure an Ethernet port on the computer to communicate with the analyzer. The IP address
n
must be 192.168.77.100. Refer to the computer operator manual for instructions on changing
the IP address.
Attach the analyzer to the Ethernet port on the computer. Use an Ethernet switch if multiple ana-
n
lyzers or Smart VacPreps will be attached.
1. Insert the installation media into the computer. The setup program automatically starts. If
the setup program does not immediately start, navigate to the media device, then locate
and double click the setup.exe file.
2. Click
Next
on the Welcome window.
3. On the Ethernet Introduction window, verify that all prerequisites have been met.
a. If not, click
not proceed with installation until these tasks have been completed.
b. If the prerequisites have been met, click Next.
Gemini VII 2390 Installation Instructions and Checklist
239-42870-02 Rev (B) — May 2021
21CFR11 E
Cancel
and complete the steps, then restart the installation program. Do

Software Installation in 21CFR11 Environments

NVIRONMENTS
8 - 5