Calibration

5-2 | Thermox CG1000 Oxygen Analyzer

Calibration vs. Verification Operations

During a verify operation, a calibration gas reading is taken, and the
difference between what the system records for the gas value and
Verify
its known value (from the gas cylinder) is recorded. Internal calibra-
tion parameters are not changed. Verify operations are often used to
check that the instrument is running within required tolerances.
During a calibration operation, software algorithms adjust the dif-
Calibration
ference between the current readings and the known calibration
gas values until they are equal.

Span vs. Zero Gas

The span gas is the high calibration gas. The zero gas is the low calibration gas.
For verify operations, only one gas is used.
NOTE

Overview

Calibrate Key
Function/Options
Initiate Cal
Start an automatic, remote, or manual calibration
Initiate Verify
Start an automatic, remote, or manual verify
Cal/Verify Data
View latest calibration and verify data
Cal Gas Values
Enter calibration gas values to match known values of
calibration gas cylinders
Configure Verify
Enter verify gas value
Enter acceptable verify range
Set Clock
Set analyzer time/date.
Define recovery duration after a cal/verify.
Table 5-1. Overview of Calibrate Key functions.
Aborting a Calibration
You can cancel a calibration or verification at any time by pressing Cancel.