7-10 | Thermox CG1000 Oxygen Analyzer
5.
Clean the old teflon tape off the fittings and place new teflon tape on
the fittings.
6.
Loop the AC harness wires around the pump inlet and outlet tubing.
This will prevent these wires from rubbing against the cover when
you reinstall the analyzer top cover.
7.
Reconnect the fittings to the pump, tightening until snug (until touch-
ing the top of the pump). Ensure that the pump is oriented correctly.
The inlet tubing (shorter length tubing) should go into the hole in the
pump with the arrow pointing towards the hole. The outlet tubing
(longer length tubing) should go into the hole with the arrow pointing
away from the hole.
8.
Place the top cover back on the analyzer.
9.
Reattach the power plug to the analyzer.
10. Turn on the analyzer.
11. Recalibrate the analyzer.