System Messages

Calibration Aborted
Calibration Required
If you see this message, select the Calibrate Key and perform a calibration. This message will not clear from the dis-
play by removing or restoring power to the control unit.
Cal Verify Aborted
Primary Calibration
System Calibrating
System Verifying Cal
6-2 | Thermox CG1000 Oxygen Analyzer
Calibration has been aborted by the user.
A regular calibration is required.
After a primary calibration is performed you must always perform a
regular calibration.
During a primary calibration the cell temperature is changed. A
system calibration cannot be performed for at least one hour. The
system calibration will determine the constants needed for the
Nernst equation.
The verification was successfully aborted by the system.
The system is performing a Primary Calibration. The message con-
tinues to be displayed during the one-hour recovery period after the
primary calibration.
System calibration is in progress.
System verification is in progress.