A-8 | Thermox CG1000 Oxygen Analyzer
Variable Name
Format
Config_flags
HEX
Iout_flags
HEX
DATA FORMAT
System configuration flags
7 – Future use
6 – Future use
5 – Future use
4 – Future use
3 – Future use
2 – Future use
1 – Future use
0 – 0 – Alarm relay deenergized
1 – Alarm relay energized
Analog outputs flags
XX XX
1
2
Byte 1 (bit #)
- Analog Output #2
7 – Future use
6 – 0 = (4 – 20) ma
1 = (0 – 20) ma
5 – 0 = holds during verify
1 = tracks during verify
4 – 0 = holds during cal
1 = tracks during cal
- Analog Output #1
3 – Future use
2 – 0 = 4 – 20 ma
1 = 0 – 20 ma
1 – 0 = holds during verify
1 = tracks during verify
0 – 0 = holds during cal
1 = tracks during cal
Byte 2 (bit #)
- Analog Output #2
7 – Future use
6 – 0 = 4 – 20 ma
1 = 0 – 20 ma
5 – 0 = holds during verify
1 = tracks during verify
4 – 0 = holds during cal
1 = tracks during cal
- Analog Output #1
3 – Future use
2 – 0 = 4 – 20 ma
1 = 0 – 20 ma
1 – 0 = holds during verify
1 = tracks during verify
0 – 0 = holds during cal
1 = tracks during cal