6.2.2 Replacing the O-ring in the oxygen outlet

Tools required:
Open-ended spanner SW 16 (only necessary if the oxygen outlet
1. If the oxygen outlet
7
cap
, remove the cap using the open-ended
47
spanner (SW 16).
2. Pull the O-ring
from the connection.
39
3. Wet a new O-ring using the oxygen lubricant
Belerub (WM 14902) and push it onto the
connection.
4. Using open-ended spanner SW 16, gently
tighten cap
.
47
Do not over-tighten

6.2.3 Dismantling the housing

Tools required:
Crosstip screwdriver, size 1,
Open-ended spanner SW 7,
Open-ended spanner SW 19 (only with AGA coupling),
Open-ended spanner SW 22,
Guide pipes, for toggle switches (from Weinmann),
Special locknut tool G 3/8 and special tool from Set WM 15348.
1. Twist the knob
as far as it will go to the left
45
so that you will have a reference point later on
when assembling.
2. Release the lid
from the knob
46
3. Hold the knob steady using the special tool to
stop it from turning, and loosen the lock screw.
Note:
Do not unscrew the screw completely; otherwise,
the knob will be dismantled into its component
parts.
4. Pull off the knob
.
45
5. If your Module is equipped with an AGA cou-
pling, loosen it with an open-ended spanner
(SW 19).
6. Unscrew union nut at inhalation
connection
.
11
7. Screw the special locknut tool onto the inhala-
tion connection
.
11
14
Repair information and repair instructions
is sealed with the
!
.
45
is sealed with the cap
7
47
O
45
46
).
47
39