Download Print this page

Agilent Technologies InfinityLab LC Series User Manual page 99

Vialsamplers
Hide thumbs Also See for InfinityLab LC Series:

Advertisement

3
Using the Module
Using the Optional Integrated Column Compartment
Method Parameters
The Method dialog box enables the user to change the method parameters
relevant for the ICC:
Valid inputs are 10 to 80 °C in increments of 0.1 °C.
NOTE
For Stoptime and Posttime, valid inputs are 0.01 to 99999 minutes in increments
NOTE
of 0.1 min.
Agilent InfinityLab LC Series Vialsamplers User Manual
Temperature:
Not Controlled: The temperature of the column oven stays uncontrolled.
At specified temperature (°C): The temperature of the column oven is
regulated to stay at the setpoint temperature specified.
Unchanged: The temperature setting stays unchanged, meaning that the
setting currently stored in the system is used by the method.
Valve Position/Column: This section provides information about the location
of the column that is being used for the given method.
Enforce column for run: Marking this checkbox restricts the execution of
the method to a specific column type, identified by its product number.
Stoptime:
As Pump/Injector: The ICC uses the same stoptime setting as specified for
the pump or the sampler.
After specified time (min): The stop time defines a time limit for the
analysis, after which all method parameters are reset to initial values.
Posttime:
Off: No post-run period is applied.
For specified time (min): After the analysis, the system stays in the
post-run state for the specified time. Using a post-time period can be
beneficial when the column needs to be re-equilibrated before starting the
next analysis (for example, gradient elutions).
99

Advertisement

loading
Need help?

Need help?

Do you have a question about the InfinityLab LC Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel