7
Error Information
Sample Cooler/Sample Thermostat Error Messages
Sample Cooler/Sample Thermostat Error
Messages
Sample temperature control voltage too low, check
fuses and wires
Error ID: 30713
The compressor voltage is below the lower threshold value.
Probable cause
1
Sample temperature control switched off due to
condensate
Error ID: 30715
The cooler/thermostat was switched off due to a condensate event.
Probable cause
1
2
Agilent InfinityLab LC Series Vialsamplers User Manual
Potential hardware error
Overfilled container
Drainage issues
Suggested actions
Please contact your Agilent service representa-
tive.
Suggested actions
Empty the condensate container. Verify that the
open end of the tubing doesn't immerse in the liq-
uid.
Verify the correct plumbing of the conden-
sate drainage system.
Make sure that no kinks or mechanical
blocks are present in the drainage system.
Avoid the formation of the siphoning effect.
Make sure that the hosting sampler is level.
213