Error Information
Sample Cooler/Sample Thermostat Error Messages
Sample Cooler/Sample Thermostat Error
Messages
Sample temperature control voltage too low, check
fuses and wires
Error ID: 30713
The compressor voltage is below the lower threshold value.
Probable cause
1
Sample temperature control switched off due to
condensate
Error ID: 30715
The cooler/thermostat was switched off due to a condensate event.
Probable cause
1
2
Agilent InfinityLab LC Series Vialsamplers User Manual
Potential hardware error
Overfilled container
Drainage issues
Suggested actions
Please contact your Agilent service representa-
tive.
Suggested actions
Empty the condensate container. Verify that the
open end of the tubing doesn't immerse in the liq-
uid.
•
Verify the correct plumbing of the conden-
sate drainage system.
•
Make sure that no kinks or mechanical
blocks are present in the drainage system.
•
Avoid the formation of the siphoning effect.
•
Make sure that the hosting sampler is level.
213