2
Site Requirements and Specifications
Specifications of the Sample Cooler
Specifications of the Sample Cooler
The Agilent Infinity II Sample Cooler is a vapor-compression refrigeration system
that uses a fluorinated greenhouse gas (HCF-134a) as the refrigerant. For
information on carbon dioxide equivalency (CDE) and global warming potential
(GWP), see the instrument label.
Table 5
Type
Weight
Dimensions (height × width × depth) 205 x 340 x 370 mm
Refrigerant gas
Supply voltage
Current
Ambient operating temperature
Ambient non-operating temperature
Humidity
Operating altitude
Safety standards:
IEC, EN, CSA, UL
ISM Classification
Agilent InfinityLab LC Series Vialsamplers User Manual
Physical Specification of the Sample Cooler
Specification
< 6 kg (< 13.2 lbs)
(8.1 x 13.4 x 14.6 inches)
HFC-134a (0.042 kg)
24 VDC
10 A max.
4 – 40 ° C (39 – 104 ° F)
-40 – 70 ° C (-40 – 158 ° F)
< 95 % r.h. at 40 °C (104 °F)
Up to 3000 m (9842 ft)
Installation category II, Pollution
degree 2
ISM Group 1 Class B
Comment
Ozone depletion poten-
tial (ODP) = 0
Non-condensing
For indoor use only
According to CISPR 11