Agilent Technologies InfinityLab LC Series User Manual page 43

Vialsamplers
Hide thumbs Also See for InfinityLab LC Series:

Advertisement

2
Site Requirements and Specifications
Performance Specifications
Table 3
Performance Specifications G7129B
Type
Specification
Carryover
<0.004 % (40 ppm) with needle
wash
Injection cycle time
18 s for draw speed 200 µL/min
Ejection speed: 200 µL/min
Injection volume: 1 µL
Minimum sample vol-
1 µL from 5 µL sample in 100 µL
ume
microvial, or 1 µL from 10 µL
sample in 300 µL microvial.
Instrument control
LC & CE Drivers A.02.12 or
above
Instrument Control Framework
(ICF) A.02.03 or above
Instant Pilot (G4208A) with firm-
ware B.02.19 or above
Lab Advisor B.02.07 or above
Communication
Controller Area Network (CAN),
Local Area Network (LAN),
ERI: ready, start, stop and
shut-down signals
Agilent InfinityLab LC Series Vialsamplers User Manual
Comments
Needle height offset has to be
adapted to ensure that needle
doesn't touch vial bottom.
Default needle height = 0
equates to 2 mm above the vial
bottom.
For details about supported
software versions refer to the
compatibility matrix of your ver-
sion of the LC & CE Drivers
Method/Conditions
Using the following conditions:
Column: Agilent Pursuit XRs
3 C18, 2.0 x 50 mm
Mobile Phase:
A: 0.1 % TFA in water
B: 0.1 % TFA in acetoni-
trile
Isocratic : %B = 40 %
Flow rate: 0.5 mL/min
Temperature: Ambient
Wavelength: 257/4 nm
Injection volume: 1 µL
Sample: 1200 ng/µL Chlor-
hexidine for UV (dissolved
with mobile phase A), 1 µL
injected and measured with
Agilent UV detector
Wash solution: H
O with
2
0.1 % TFA (3 s)
43