4
Preparing the Module
Setting up the Vialsampler
Table 16
Control settings
Figure 30
Control settings
Agilent InfinityLab LC Series Vialsamplers User Manual
The Sampler control parameters are in the following sections:
•
Missing Vial
Mark the Ignore missing vial check box to specify that, if a
vial is missing, the injector ignores it and continues with a
6-second dummy run. The message "Missing vial <x>" is
logged, and the system continues with the next injection.
•
Illumination
Toggles the illumination of the sample area, On or Off.
•
At Power On
The section is available when a cooler/thermostat is
installed and configured. Mark the Turn on Thermostat
check box to specify that the cooler/thermostat is
switched on automatically when the instrument is
switched on.
•
Thermostat
The section is available when a cooler/thermostat is
installed and configured and the Constant temperature
mode is selected.
Select On to switch on the cooler/thermostat. Specify the
required temperature in the adjacent field. The specified
temperature must be at least 5 °C below ambient for
proper temperature control.
Select Off to switch off the cooler/thermostat.
•
Automatic Turn On
You can set a date and time at which the cooler/thermo-
stat switches on automatically.
•
Pump connected to Sampler
Use this section to specify the pump that is used with the
Vialsampler. If more than one pump is configured, display
the drop-down list and select the appropriate pump from
the list.
•
Enable Analysis
This feature requires LC & CE Drivers A.02.19 or newer and
is only available for the Sample Thermostat. With this func-
tion, you can specify if the analyses should start With any
temperature or only when the Temperature is within ± 2 °C
range of the setpoint temperature.
NOTE
The Enable Analysis section is disabled when Not con-
trolled is selected in the Temperature section.