Download Print this page

Agilent Technologies InfinityLab LC Series User Manual page 241

Vialsamplers
Hide thumbs Also See for InfinityLab LC Series:

Advertisement

8
Maintenance
Exchange the Sample Loop Assembly
Exchange the Sample Loop Assembly
When
Tools required
p/n
8710-1924
Parts required
#
1
OR
1
OR
1
Preparations
Risk of injury by uncovered needle
WAR N IN G
An uncovered needle is a risk of harm to the operator.
Do not open the safety cover of the needle station during normal
operation.
Wear safety goggles and safety gloves when removing the needle
assembly.
When the instrument setup has changed, configure the new setup in the online
NOTE
view of the Chromatographic Data System. See
Agilent InfinityLab LC Series Vialsamplers User Manual
The sample loop is visibly damaged.
Leaks or blockages are observed.
Description
Open-end wrench 14 mm
p/n
Description
G7129-60500
Sample Loop 100 µL
G7129-60400
Sample Loop 40 µL
G7129-60300
Sample Loop 20 µL
Finish any pending acquisition job.
Stop the flow at the pump and remove the solvent lines from the eluent bottles to avoid spilling
solvent.
Close the shutoff valves at the pump if available.
Table 15
on page 155.
241

Advertisement

loading
Need help?

Need help?

Do you have a question about the InfinityLab LC Series and is the answer not in the manual?

Questions and answers