7
Error Information
Vialsampler Error Messages
Metering home failed
Error ID: 34224
The home position of the metering plunger is not found or the plunger is unable
to move.
Probable cause
1
2
Automatic Referencing procedure failed
Error ID: 34225
The execution of the Automatic Referencing procedure was aborted by the
system due to an unexpected error.
Probable cause
1
2
Agilent InfinityLab LC Series Vialsamplers User Manual
Blockage in the flow path.
Potential hardware error
Potential hardware error
Physical blockage
Suggested actions
Check if there is a significant pressure difference
(> 10 – 15 bar) between the mainpass and the
bypass position of the injection valve. If yes, sys-
tematically replace the components of the main-
pass: Needle seat, Needle, Rotor seal, Sample
loop
Please contact your Agilent service representa-
tive.
Suggested actions
Please contact your Agilent service representa-
tive.
Look for potential sources of physical blockage
in the path of the gripper arm and eliminate
them. Restart the sampler and observe if the
problem persists.
202
Need help?
Do you have a question about the InfinityLab LC Series and is the answer not in the manual?
Questions and answers