Dräger Carina Instructions For Use Manual page 61

Sub-acute care ventilator
Hide thumbs Also See for Carina:
Table of Contents

Advertisement

Test step
6 Check the Airway pres-
sure low !!! alarm
7 Check apnea ventilation
After a successful device check, Carina is ready for use with the patient.
After the device check
Switch Carina to standby mode, see page 77.
Instructions for use Carina SW 3.2n
Measure
Confirm the Airway pressure
high !!! alarm.
Disconnect the test lung from
the breathing circuit.
Connect the test lung to the
breathing circuit.
Set the ventilation mode:
SPN-CPAP/PS
Setting alarm limits:
Paw high:
45 mbar
Tapn:
5 s
Setting ventilation parameters:
Apn.Vent.:
On
Pmax:
40 mbar
VTapn:
500 mL
Simulate spontaneous breath-
ing by squeezing the test lung.
Then stop the breathing.
Requirement
Optical alarm signals:
The Airway pressure low
!!! alarm message is dis-
played and the Audio
paused 2 min. alarm LED
flashes red.
Acoustic alarm signal: High-
priority alarm signal
After the apnea alarm time
has elapsed:
Optical alarm signals:
The Apnea Ventilation !
alarm message is displayed
and the Audio paused
2 min. alarm LED lights yel-
low.
Acoustic alarm signal: Low-
priority alarm signal
Getting started
Result
61