Operation

Safety information on ventilation

WARNING
Risk of patient injury
A component of the volume supplied by the
device remains in the breathing circuit during
ventilation. This component depends on the
ratio of hose compliance to lung compliance.
The device does not take account of the com-
pliance of the breathing circuit.
The volume losses must be manually cor-
rected. For additional information, see chapter
"Volume losses during ventilation"
on page 144.
WARNING
Risk of increased CO
The risk of dead space ventilation is increased
when setting small volumes.
Observe the patient very closely.
WARNING
Risk of insufficient ventilation
Leakages may cause the applied volume to be
less that the set volume.
Prevent leakages.
WARNING
Risk of patient injury
Critical airway pressures may result with Vol-
ume Guarantee.
Set Pmax accordingly.
CAUTION
The Apnea !!! alarm is only displayed in an apnea
situation if apnea monitoring is activated.
Set the apnea alarm time Tapn.
66
rebreathing
2
Instructions for use Carina SW 3.2n