Alarms

Suppressing the acoustic alarm signal and the nurse call

The following actions are possible with the Audio
paused 2 min. key:
– Suppressing the acoustic alarm signal for a
maximum of 2 minutes
– Suppressing alarm generation by the nurse
call for a maximum of 2 minutes
If an alarm with the same or higher priority arises
during this time, the acoustic alarm signal sounds
once. Alarm generation by the nurse call will not be
activated.
If the cause of the alarm is not rectified after a max-
imum of 2 minutes:
– The acoustic alarm signal is initiated again.
– Alarm generation by the nurse call is initiated
again.
NOTE
For the Airway pressure low !!! alarm, the
acoustic alarm signal and alarm generation by the
nurse call can be suppressed for a maximum of
1 minute only.
 Press the Audio paused 2 min. (A) key.
The acoustic alarm signal and alarm generation by
the nurse call are suppressed for a maximum of
2 minutes.
Inlet filter missing !!
B
The device displays the
alarm line.
86
A
(B) symbol in the
To reactivate the alarm signal and alarm generation
by the nurse call before the suppression time has
elapsed:
 Press the Audio paused 2 min. (A) key again.
Instructions for use Carina SW 3.2n