Operation
WARNING
Missing O
supply
2
Patients who require an increased O
tration are at risk in the event of failure of the
oxygen source.
Make sure there is an emergency oxygen sup-
ply, e.g., O
from compressed gas cylinders.
2
CAUTION
Insufficient O
supply
2
If the O
supply is insufficient (e.g., because of a
2
kinked LPO hose), the patient will not receive the
increased O
concentration from the LPO supply.
2
In this case an alarm is not triggered.
In the case of patients who require an increased
O
concentration use external O
2
WARNING
Risk of patient injury
The patient will be at risk if unapproved hose
systems are used between the device and the
oxygen supply.
Only use hoses approved for medical use and
for use with oxygen.
NOTE
The automatic O
mixing of the device does not
2
operate in LPO mode. The O
only be set by the following parameters:
– For leakage valve: as a function of Pmean and
LPO flow
– For expiratory valve: as a function of minute
volume and LPO flow
NOTE
If O
is supplied simultaneously from the central
2
gas supply or gas cylinder and from the O
centrator the LPO feed will be blocked.
 Observe the instructions for use of the oxygen
source used, e.g., the O
74
concen-
2
monitoring.
2
concentration can
2
con-
2
concentrator.
2
Connecting the O
2
Carina
A B
1 Remove the protective cap (A) from the LPO
connection (B).
2 Connect the O
supply hose (C) to the LPO
2
connection.
3 Connect the O
supply hose to the oxygen
2
source (D). Observe the instructions for use for
the O
concentrator.
2
Instructions for use Carina SW 3.2n
concentrator to
D
C