Setting Ranges For Ventilation - Dräger Carina Instructions For Use Manual

Sub-acute care ventilator
Hide thumbs Also See for Carina:
Table of Contents

Advertisement

Operation

Setting ranges for ventilation

Ventilation
parameter
VC-SIMV
1)
VT
[mL]
100 to 2000
2)
Pinsp
[mbar,
cm H
O]
2
PEEP [mbar,
3 to 20
cm H
O]
2
3)
FiO
[%]
21 to 100
2
Trigger
Off, Normal,
Sensitiv.
RR [/min]
5 to 50
4)
Ti
[s]
0.3 to 8
or I:E
1:39 to 2:1
Ramp
Auto/0.1 to 2.0
5)
Pmax
[mbar,
LeakV: 5 to 40
cm H
O]
ExpV: 5 to 50
2
6)
ΔPsupp
LeakV: 0 to 37
[mbar, cm H
O]
ExpV: 0 to 47
2
Timax [s]
Apn.Vent.
7)
RRapn
[/min]
7)
VTapn
[mL]
VG
8)
Aut.Adapt
9)
Tadpt
[min]
9)
PEEPadp
[mbar, cm H
O]
2
9)
ΔPSadpt
[mbar, cm H
O]
2
1) For SPN-CPAP/PS the following applies: Only with activated Volume Guarantee VG.
2) For PC-BIPAP and PC-AC the following applies: Only Pinsp values that are at least 2 mbar above the set PEEP value can
be set.
3) Only with HPO O
supply.
2
4) For VC-SIMV/PS and PC-BIPAP/PS the following applies: The maximum inspiratory time Timax for pressure-supported
breaths is derived from Ti, see page 150.
5) For SPN-CPAP/PS (VG) the following applies: Only with activated apnea ventilation or activated Volume Guarantee VG.
6) For SPN-CPAP/PS the following applies: Only with deactivated Volume Guarantee VG. For PC-BIPAP and VC-SIMV the
following applies: only with trigger activated
7) Only with apnea ventilation activated.
8) Only in NIV application mode with Volume Guarantee VG deactivated.
9) Only with Aut.Adapt function activated.
68
VC-AC
100 to 2000
3 to 20
21 to 100
Off, Normal,
Sensitiv.
5 to 50
0.3 to 8
1:39 to 2:1
Auto/0.1 to 2.0
LeakV: 5 to 40
ExpV: 5 to 50
Ventilation mode
PC-BIPAP
PC-AC
LeakV: 5 to 40
LeakV: 5 to 40
ExpV: 5 to 50
ExpV: 5 to 50
3 to 20
3 to 20
21 to 100
21 to 100
Off, Normal,
Off, Normal,
Sensitiv.
Sensitiv.
5 to 50
5 to 50
0.3 to 8
0.3 to 8
1:39 to 2:1
1:39 to 2:1
Auto/0.1 to 2.0
Auto/0.1 to 2.0
LeakV: 0 to 37
ExpV: 0 to 47
Instructions for use Carina SW 3.2n
SPN-CPAP/PS
(VG)
100 to 2000
3 to 20
21 to 100
Normal, Sensitiv.
Auto/0.1 to 2.0
LeakV: 5 to 40
ExpV: 5 to 50
LeakV: 0 to 37
ExpV: 0 to 47
0.4 to 10.0
On, Off
5 to 50
100 to 2000
On, Off
On, Off
5 to 120
3 to 20
0 to 37