Intrahospital Patient Transport; Safety Information For Intrahospital Patient; Transport; Increasing The Toppling Stability - Dräger Carina Instructions For Use Manual

Sub-acute care ventilator
Hide thumbs Also See for Carina:
Table of Contents

Advertisement

Assembly and preparation

Intrahospital patient transport

Safety information for intrahospital pa-
tient transport
WARNING
Risk of patient injury
Changes to the patient's condition or damage
to the device during transport endanger the
patient.
The patient must be monitored continuously
by users.
WARNING
Risk of patient injury
If the internal battery is discharged and the ex-
ternal power supply fails, the device cannot
ventilate.
Only start intrahospital patient transport if the
internal battery is sufficiently charged.
CAUTION
Risk of damage to device or personal injury
If Carina is placed on the bed when transporting
patients, the device may fall down.
Do not place the device on the patient's bed.
CAUTION
Risk of toppling over
The medical device may topple over if incorrectly
handled.
Do not tilt the trolley more than 10°.
Hold the device firmly by the handle to move or
position it.
Move the trolley at walking speed only. The dan-
ger of toppling over is greater at thresholds, on
uneven surfaces and ramps. Reduce speed even
more.
54

Increasing the toppling stability

To ensure that the device cannot topple over, the
accessories must be moved to the most advanta-
geous position:
– Set hinged arm to minimum extension.
– Attach hoses and cables as close as possible to
the trolley.
– Attach breathing gas humidifier to the trolley,
not the device.
Instructions for use Carina SW 3.2n