Starting Ventilation; Auto Wake-Up - Dräger Carina Instructions For Use Manual

Sub-acute care ventilator
Hide thumbs Also See for Carina:
Table of Contents

Advertisement

Starting ventilation

Before using on the patient:
1 Check readiness for operation, see page 59.
2 Sett ventilation modes and ventilation parame-
ters, see page 67.
3 Setting the alarm limits, see page 84.
A
 Press the Start/Standby (A) key.
The start of ventilation is indicated by a single tone.
Instructions for use Carina SW 3.2n

Auto wake-up

1 Connect the patient.
2 Carina starts ventilation on detection of the in-
spiratory effort with the previously set ventila-
tion settings.
3 Monitor the start of ventilation. Check the venti-
lation settings and adjust if necessary.
WARNING
Risk of patient injury
After connecting the patient, monitor the start
of ventilation. If the patient's inspiratory effort
is not sufficient, the device cannot start venti-
lation.
Ventilation must then be started manually.
Getting started
63

Advertisement

Table of Contents
loading

Table of Contents