Setting the alarm volume

WARNING
Failure to hear alarm signals in a noisy envi-
ronment
Alarm situations will not be noticed.
Set the volume of alarm signals so that they
can be perceived.
The minimum alarm volume can be configured by
service personnel and experts in the Service > Bi-
omed menu.

Setting the acoustic alarm signal

The acoustic alarm signal (Standard or Dräger) can
be configured by service personnel and experts in
the Service > Biomed menu.

Selecting the screen display

Prerequisite: The Config. menu is opened.
1 Press the Screen key.
2 Select Curves or Measure by turning the rotary
knob and push to confirm.
The current setting is shown in the Screen key.
Instructions for use Carina SW 3.2n
Prerequisite: The Config. menu is opened.
1 Press the Volume key.
2 Set the value from 1 to 5 by turning the rotary
knob and push to confirm.
The specific alarm volume is set.
NOTE
When setting the alarm volume (Volume), the
user will no longer be able to set the level below
the minimum volume set in the Service > Biomed
menu.
Configuration
91