Monitoring Functions - Dräger Carina Instructions For Use Manual

Sub-acute care ventilator
Hide thumbs Also See for Carina:
Table of Contents

Advertisement

Monitoring functions

Sound pressure level L
measured according to IEC60601-1-8 and
A1:2012:
Defined user's operating location: Front of device
at 1 m (39 in) distance and 1.5 m (59 in) height.
Acoustic alarm signal, standard setting (according
to IEC 60601-1-8)
Range for high-priority alarms
Range for medium-priority alarms
Range for low-priority alarms
Step size
Acoustic alarm signal/ Dräger setting (common
acoustic alarm signal for Dräger ventilators)
Range for high-priority alarms
Range for medium-priority alarms
Range for low-priority alarms
Step size
Power supply failure alarm and auxiliary alarm
Inspiratory minute volume
Alarm for upper alarm limit
Range
Resolution
Alarm for lower alarm limit
Range
Resolution
Airway pressure
Alarm for upper alarm limit
Range
Resolution
Delay time to alarm Airway pressure low !!!
Non-invasive ventilation (NIV)
Invasive ventilation (Tube)
Respiratory rate
Alarm
Range
Resolution
Instructions for use Carina SW 3.2n
of the alarm signals,
PA
approx. 73 dB (A) to approx. 78 dB (A)
approx. 68 dB (A) to approx. 76 dB (A)
approx. 53 dB (A) to approx. 73 dB (A)
Can be set in 5 steps
approx. 67 dB (A) to approx. 81 dB (A)
approx. 68 dB (A) to approx. 83 dB (A)
approx. 65 dB (A) to approx. 78 dB (A)
Can be set in 5 steps
approx. 60 dB (A)
MV
If the upper alarm limit has been exceeded
2 to 60 L/min
0.1 L/min
If the lower alarm limit has been exceeded
0.1 to 39 L/min
0.1 L/min
Paw
If the alarm limit has been exceeded
10 to 55 mbar (or hPa or cm H
1 mbar (or hPa or cm H
Tdiscon
0, 15, 30, 60, 90, 120 s
0, 15, 30, 60 s
RR high
If the alarm limit has been exceeded for one minute
10 to 74/min, Off
1/min
Technical data
O)
2
O)
2
129