EMC declaration

General information

The device was tested for electromagnetic compat-
ibility using accessories from the list of accessories.
Other accessories may only be used if they do not
compromise the electromagnetic compatibility. The
use of non-compliant accessories may result in in-
creased electromagnetic emissions or decreased
electromagnetic immunity of the device.

Electromagnetic environment

The emissions from this device were tested in the
following frequency ranges:
Emissions
Conducted emissions
Radiated emissions
Harmonic emissions
Voltage fluctuation emissions (flicker)

Electromagnetic immunity

This device may only be used in environments
specified in section "Intended use" on page 16.
Immunity against
Electrostatic discharge (ESD) (IEC 61000-4-2)
Fast transient electrical disturbances (bursts)
(IEC 61000-4-4)
Impulse voltages (surges) (IEC 61000-4-5)
Magnetic fields at a mains frequency of 50 Hz
(IEC 61000-4-8)
Instructions for use Carina SW 3.2n
The device may be used in the direct vicinity of oth-
er devices only if Dräger has authorized this device
arrangement. If no authorization has been given by
Dräger, it must be ensured that the device functions
correctly in the desired arrangement before using it.
The instructions for use for the other devices must
be followed.
Compliance
Class B, group 1 (150 kHz to 30 MHz)
Class B, group 1 (30 MHz to 1 GHz)
Requirement is met
Requirement is met
Test level
Contact discharge: ±8 kV
Air discharge: ±15 kV
Power cable: ±2 kV
Longer input/output lines: ±1 kV
Voltage, external conductor – external conductor:
±1 kV
Voltage, external conductor – protective ground con-
ductor: ±2 kV
30 A/m
Technical data
137