Additional messages on the screen

Message
Battery indication not relia-
ble !
Standby activated
Key lock activated

Emergency ventilation

WARNING
Risk of patient injury
If a fault is detected in the medical device, its
life-support functions may no longer be as-
sured.
Ventilation of the patient using an independ-
ent ventilation device must be started without
delay, if necessary with PEEP and/or an in-
creased inspiratory O
with a manual resuscitator).
In the event of a technical fault, e.g., a pressure
sensor fault, the device switches to emergency
ventilation at 21 Vol% FiO
controlled ventilation at the required ventilation
pressure even in the event of sensor faults.
Emergency ventilation is a pressure-controlled ven-
tilation with reduced quality. The trigger function is
deactivated. Volume Guarantee is no longer avail-
able. Functions are not monitored.
Instructions for use Carina SW 3.2n
Cause
Battery display is inaccurate.
The device is in standby mode,
there is no ventilation.
Locked function is active.
concentration e.g.,
2
. The turbine maintains
2
Problem solving
Remedy
Improve the accuracy,
see page 79.
If there is no improvement, inform
DrägerService.
Start ventilation: Press the
Start/Standby key.
Press the Locked key again.
103