Connecting the breathing gas humidifier

Safety information for connecting the
breathing gas humidifier
WARNING
Risk of patient injury
When using a breathing gas humidifier with-
out heated hoses condensate may increase
the resistance.
Use a water trap.
WARNING
Risk of damange to the device and of patient
injury
Position device and breathing gas humidifier
on a flat, stable surface. Otherwise the breath-
ing gas humidifier may fall down.
WARNING
Risk of patient injury
Increased resistance values may cause hy-
poventilation.
When using a breathing gas humidifier do not
use an additional heat and moisture ex-
changer (HME).
Instructions for use Carina SW 3.2n
Tube
PC-AC

Fixing the hinged arm

To prevent the breathing hose from kinking, Dräger
recommends the use of a hinged arm.
 Attach the hinged arm with rail clamp to the uni-
versal rail on the trolley and fix in position.
WARNING
Risk of patient injury
When using active humidification the device
may not detect a disconnection due to the re-
stricted monitoring function.
Use suitable monitoring.