Breathing gas humidifier without inspir-
atory hose heater
For breathing circuits with leakage valve
1 Set the switch to leakage valve
see page 42.
A
B
C
2 Fit the bacterial filter (A) to the inspiratory port
(B).
3 Connect the breathing hose (C) to the bacterial
filter and the breathing gas humidifier (D).
4 Connect the breathing hose (E) to the breathing
gas humidifier (D) and the water trap (F). Con-
nect another breathing hose (E) to the water
trap.
5 Position the water trap (F) vertically at the low-
est point of the hose connection.
6 Connect the leakage valve (G). Observe the di-
rection.
Instructions for use Carina SW 3.2n
,
E
E
F
D
For breathing circuits with expiratory valve
1 Set the switch to expiratory valve
see page 42.
H
A
B
C
I
G
2 Fit the bacterial filter (A) to the inspiratory port
(B).
3 Connect the breathing hose (C) to the bacterial
filter and the breathing gas humidifier (D).
4 Connect the breathing hose (E) to the breathing
gas humidifier (D) and the water trap (F). Con-
nect another breathing hose (E) to the water
trap.
5 Position the water trap (F) vertically at the low-
est point of the hose connection.
6 Connect the expiratory valve (G). Observe the
direction.
7 Connect the pilot line (H) to the plug
Assembly and preparation
,
E
E
F
D
(I).
G