Technical data
Displayed measured values (cont.)
Respiratory rate
Range
Resolution
Accuracy
T
0...90
Mean airway pressure
Range
Resolution
Accuracy
Minute volume
Range
Resolution
Accuracy
T
0...90
Leakage minute volume
Range
Resolution
Accuracy
T
0...90
Adjustment time in AutoAdapt
Range
Resolution
Accuracy
Airway pressure waveform display
Range
Flow waveform display
Range
Airway pressure bar graph
Range
128
RR
0 to 75/min
1/min
±1/min
approx. 60 seconds
Pmean
0 to 68 mbar (or hPa or cm H
0.1 mbar (or hPa or cm H
±(8 % of measured value +1.2 mbar) when using
MP00312, MP00313 breathing circuits (without
leakages)
MVi, MV
0 to 60 L/min, BTPS
0.1 L/min
±10 % of measured value or ±0.6 L/min, whichever
is greater (without leakage, and taking account of
hose compliance correction, "Volume losses during
ventilation" on page 144)
approx. 35 s
MVleak
0 to 99 L/min, BTPS
0.1 L/min
±10 % of measured value or ±0.6 L/min, whichever
is greater (without leakage, and taking account of
hose compliance correction, "Volume losses during
ventilation" on page 144)
Approx. 35 s
Tadpt
0 to 120 min
1 min
1 min
Pressure
0 to 80 mbar (or hPa or cm H
Flow
–160 to +160 L/min
Pressure
0 to 60 mbar (or hPa or cm H
Instructions for use Carina SW 3.2n
O)
2
O)
2
O)
2
O)
2

Advertisement