For your safety and that of your patients

Product-specific safety information

WARNING
Risk of patient injury
The device may not be used for the acute inpa-
tient treatment of unstable patients that have
to be ventilated with extensive, continuous
monitoring in accordance with IEC 60601-2-12
or ISO 80601-2-12.
WARNING
Risk of incorrect use
This medical device is only intended to be
used by the target group "users".
WARNING
Risk of malfunctions
Prohibited modifications to the medical de-
vice lead to malfunctions.
This medical device must not be modified
without permission from the manufacturer.
WARNING
Risk of patient injury
Penetrating liquid may cause malfunction of
or damage to the device, which may endanger
the patient.
Do not place any containers with liquid on or
above the device.
Make sure that no liquids penetrate the device
during surface disinfection.
WARNING
Risk of fire
Do not use the medical device in conjunction
with flammable gases or flammable solutions
that can mix with air, oxygen or nitrous oxide,
or other sources of ignition since the medical
device could ignite.
Do not allow the medical device to come into
contact with sources of ignition.
12
WARNING
Risk of fire
If medications or other substances based on
flammable solvents such as, e.g., alcohol,
reach the breathing circuit, there is a risk of
fire.
If highly flammable substances are used for
disinfection, adequate airing must be en-
sured.
WARNING
Risk of fire
Due to oxygen enrichment in the ambient air,
the medical device can ignite. Medical device
malfunctions can increase the O
tion in the ambient air.
Only use the medical device in adequately
ventilated rooms.
WARNING
Risk of fire
Do not use a defibrillator at the same time as
additional oxygen is administered.
WARNING
Risk of patient injury
Magnetic fields may affect the correct opera-
tion of the medical device.
Do not use the medical device in the vicinity of
nuclear magnetic resonance scanners (MRI,
NMR, NMI).
WARNING
Risk of patient injury
Hyperbaric chambers may impair correct
functioning of the medical device.
Do not use the medical device in hyperbaric
chambers.
Instructions for use Carina SW 3.2n
concentra-
2