Alarm Delays - Dräger Carina Instructions For Use Manual

Sub-acute care ventilator
Hide thumbs Also See for Carina:
Table of Contents

Advertisement

Technical data

Alarm delays

Alarms are not displayed immediately when the
alarm limit is exceeded.
Alarm
Airway pressure high !!!
Airway pressure low !!!
Apnea !!!
Apnea Ventilation !
Breath. circuit incorrect !!!
Continuous Paw high !!!
Continuous Paw low !!!
HPO supply insufficient !!!
Large leakage !
LPO pressure high !!
MV low !!!
RR high !!!
136
Delay time
If the alarm limit is exceeded in 2 consecutive breathing phases
>Tdiscon + 5 s
After Tapn has elapsed, if a time has been set for Tapn
After Tapn has elapsed, if a time has been set for Tapn
>37 s
>15 s
>30 s (at least 3 breaths)
>30 s
If the permissible patient leakage is exceeded from the 11th breathing
phase
If the alarm limit is exceeded in 2 consecutive breathing phases
60 s at start of ventilation and after reset of the alarm
Airway pressure low !!!
60 s
Instructions for use Carina SW 3.2n