Dräger Carina Instructions For Use Manual page 131

Sub-acute care ventilator
Hide thumbs Also See for Carina:
Table of Contents

Advertisement

Operating data characteristics (cont.)
Dew point
Oil content
Particle size
LPO
O
operating pressure
2
O
flow
2
Connector hose
O
humidity
2
Noise emission of device during ventilation at typi-
cal setting (mean sound pressure level Leq (A))
(see page 133)
Dimensions of basic device (width x height x
depth)
Weight of basic device
Heating of breathing gases in device
Screen
Device outputs
Digital output
MEDIBUS protocol
Baud rate
Data bits
Parity
Stop bit
MEDIBUS.X protocol
Baud rate
Data bits
Parity
Stop bit
Pin assignment
Pin 2
Pin 3
Pin 5
Instructions for use Carina SW 3.2n
5 K below ambient temperature
3
<0.1 mg/m
Dust-free O
flow (filtered with pore size <1 µm)
2
0 bar to 0.5 bar
0 kPa to 50 kPa
0 psi to 7.25 psi
0 to 10 L/min
Max. ∅ 7 mm (0.27 in)
Without condensation
40 dB (A)
175 x 275 x 385 mm (6.9 x 10.8 x 15.2 in)
5.5 kg (12.1 lbs)
Typically 2 K, max. 5 K
TFT color screen 14.5 cm (5.7 in)
RS232 COM port, interface for MEDIBUS or
MEDIBUS.X protocol
9600 baud
8
None
1
19200 baud
8
Even
1
RXD
TXD
GND
Technical data