Configuration
Setting the language
The language can only be set in standby mode.
 Select the line Language by turning the rotary
knob and push to confirm.
 Select the desired language by turning the ro-
tary knob and push to confirm.
Setting the software protocol
The software protocol for data transfer can only
be set in standby mode.
 Select the line Protocol with the rotary knob.
 Set MEDIBUS or MEDIBUS.X by turning the ro-
tary knob and push to confirm.
Configuring the flow reduction (Anti Air
Shower)
The function can only be set in standby mode.
 Select the line Anti Air Shower by turning the
rotary knob and push to confirm.
 Set On or Off by turning the rotary knob and
push to confirm.

Displaying the contact information

Prerequisite: The Service menu is open.
 Press the Contact key.
The following information for contacting
DrägerService is displayed:
– Organization
– Telephone number
– Customer ID
– City
90

Displaying the logbook

Prerequisite: The Service menu is open.
 Press the Logbook key.
The logbook shows the technical alarms as infor-
mation for the service. Technical alarm messages
are only available in English. Only the technical
alarms from the last 48 hours are displayed on the
device. The complete logbook can be downloaded
or reviewed by specialized service personnel.
Once there are more than approx. 500 logbook en-
tries, the oldest entries will be overwritten.
The entries in the logbook are preserved even after
the device is switched off and on again, and after a
power failure of any duration.
Instructions for use Carina SW 3.2n