Alarm message
Invalid date !
Key overused !!!
(Key overused !!!)
Large leakage !
Leakage valve blocked !!!
LPO pressure high !!
MEDIBUS !
(MEDIBUS !)
MV high !!
MV low !!!
O2 calibration failed !
(O2 calibration failed !)
Instructions for use Carina SW 3.2n
Internal
Cause
priority
number
460
The date setting is incorrect.
175
Key pressed too often.
452
The breathing circuit or mask
is leaking. Note: Only dis-
played after the 3rd breath.
265
Low flow because of deposits
in leakage valve, Risk of CO
rebreathing.
410
LPO is set although O
pressed gas supply (HPO) is
connected.
462
An unexpectedly large
amount of data is being re-
ceived over the RS232 COM
port.
414
The upper alarm limit of the
minute volume has been ex-
ceeded.
Leakage in breathing system Make sure that the breathing
Device malfunction
209
The lower alarm limit of the
minute volume has been ex-
ceeded.
Device malfunction
458
Calibration of O
faulty.
Remedy
Confirm the alarm message by
pressing the rotary knob. Cor-
rect date. Inform
DrägerService.
Check the function of the keys.
Check mask, filters, and other
components for leakages and
loose connections.
Replace leakage valve.
2
com-
Disconnect O
2
gas supply (HPO) or set HPO.
Check the external device on
the RS232 COM port.
Check patient condition, check
breathing pattern, correct
alarm limit if necessary.
system is leakage-free.
Disconnect the patient from
the device and immediately re-
sume ventilation using an-
other, independent ventilator.
Inform DrägerService.
Check patient condition, check
breathing pattern, correct
alarm limit if necessary.
Disconnect the patient from
the device and immediately re-
sume ventilation using an-
other, independent ventilator.
Inform DrägerService.
dosing is
Inform DrägerService.
2
Problem solving
compressed
2
101