SONET_Usr.bk Page 191 Wednesday, December 13, 2000 11:11 AM
Description
HOW TO:
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Selecting and Using "OTHER" Features

REI-L/MS-REI Result/Enable

REI-L/MS-REI Result/Enable

Before running a test, you can choose to enable or disable the SDH measurements of
MS-REI and MS-AIS or the equivalent SONET measurements of REI-L/AIS-L .
This function is available on the
Choose MS-REI/MS-AIS or REI-L/AIS-L ENABLE
When set to OFF, the MS-REI and REI-L measurements are suppressed.
Cumulative, Short Term and Analysis measurements are affected alike.
The MS-REI measurement is twinned with B2-BIP with regard to the G.826 PUAS
measurement. This results in the B2-BIP PUAS measurement showing N/A when
MS-REI is set to OFF.
OTHER
MISCELLANEOUS
or
ON
display.
as required.
OFF
191