Agilent Technologies OmniBER 720 User Manual page 86

Hide thumbs Also See for OmniBER 720:
Table of Contents

Advertisement

SONET_Usr.bk Page 83 Wednesday, December 13, 2000 11:11 AM
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Selecting Test Features - SONET Operation
Using Receive Overhead Capture
- captures activity after the first occurrence of a deviation from your
ON NOT
specified overhead state. Can be used for transient detection from a specified
expected state.
5 Up to 16 records of overhead state are provided. Each record will represent
between 1 and 64,000 frames. A capture is started by pressing CAPTURE
and terminates when up to 16 records have been captured. The capture
START
can be terminated earlier by pressing CAPTURE
.
STOP
83