SONET_Usr.bk Page 97 Wednesday, December 13, 2000 11:11 AM
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Selecting Test Features - SONET Operation
Stressing Optical Clock Recovery Circuits
Choose the required STRESSING PATTERN.
The G.958 test pattern consists of 7 consecutive blocks of data as follows:
the first row of section overhead bytes, ALL ONES, a PRBS, the first row of
section overhead bytes, ALL ZEROS, a PRBS and the first row of section
overhead bytes.
2
If you choose ALL ONES or ALL ZEROS as the stressing pattern, choose the
number of bytes in the BLOCK LENGTH.