Generating Overhead Sequences - Agilent Technologies OmniBER 720 User Manual

Hide thumbs Also See for OmniBER 720:
Table of Contents

Advertisement

SONET_Usr.bk Page 38 Wednesday, December 13, 2000 11:11 AM
Selecting Test Features - SDH Operation

Generating Overhead Sequences

Generating Overhead Sequences

Description
You may insert a sequence of patterns into a functional group of overhead bytes for
testing or troubleshooting purposes.
1 Set up the SDH transmit interface and payload required. See "Setting Transmit
HOW TO:
2 Select
3 Choose the type of sequence required.
4 Choose the overhead type as required.
5 Choose the byte or bytes of overhead required.
6
38
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Interface" page 20.
, SDH, SEQUENCES as shown above.
TEST FUNCTION
SINGLE RUN - runs the sequence once and then stops.
REPEAT RUN - runs the sequence repeatedly until STOPPED is chosen.
RSOH- Regenerator Section Overhead
MSOH - Line Overhead
POH - Path Overhead
Set up the required number of data patterns and the number of frames in which
each data pattern should appear.
Your sequence is derived from up to 5 blocks of hexadecimal data. Each block
can be transmitted in up to 64,000 frames.
The data and the number of frames are set using
INCREASE DIGIT
DECREASE DIGIT
.

Advertisement