Selecting Test Features - SONET Operation
Using Receive Overhead Monitor
4 If POH (Path Overhead) is chosen, choose the source of the overhead, SPE or
VTSPE.
J1 and J2 bytes can be monitored under Path Overhead or Trace Messages
5 If TRACE MESSAGES is chosen, you can monitor a data message to verify
portions of the network.
If the 16 byte CRC7 message structure is detected, the 15 characters within the
message are displayed.
If the CRC7 structure is not detected in J1, the 64 byte message format is assumed
and displayed.
If the CRC7 structure is not detected for J0 or J2, all 16 bytes are displayed.
6 If LABELS is chosen, the S1 sync status, STS path label (C2) and the VT Path
label (V5) are monitored.
7 If APS MESSAGES is chosen, choose the TOPOLOGY,
(GR-1230). The K1 and K2 bits are monitored.
RING
If any abnormal behavior is observed on a particular path or section overhead byte,
or an associated group of bytes (3XA1,3XA2; D1 - D3), the
display of
TEST FUNCTION
on the suspect byte or bytes on a frame by frame basis. See "Using Receive
Overhead Capture " page 82.
can be used to "Zoom" in
OVERHEAD CAPTURE
(GR-253) or
LINEAR
RECEIVE
77